FDA (Devices) recall action 95331

Boston Scientific Corporation recalled 22 products on 2024-10-23

The FDA (Devices) publishes a recall like this one as 22 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 22 pages.

Products
22
Announced
2024-10-23
Critical
0
Units
790

Why these products were recalled

Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a refere…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–22 of 22

  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific Recalls AVVIGO+ Guidance System Due to Software Anomaly

    Boston Scientific is recalling 4 units of the AVVIGO+ Multi-Modality Guidance System because a software error displays the wrong grid overlay during Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific Recalls AVVIGO+ Guidance Systems Due to Software Display Error

    Boston Scientific is recalling 40 AVVIGO+ Multi-Modality Guidance Systems because a software anomaly causes an incorrect grid overlay to display during Live or Record mode when used with specific catheters.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities whi
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific Recalls AVVIGO+ Guidance System Due to Software Anomaly

    Boston Scientific is recalling 14 units of the AVVIGO+ Multi-Modality Guidance System due to a software anomaly that displays an incorrect grid overlay during Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities whi
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific Recalls AVVIGO+ Guidance System Due to Software Anomaly

    Boston Scientific is recalling 15 units of the AVVIGO+ Multi-Modality Guidance System because a software anomaly causes an incorrect grid overlay to display during Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM EU + ROW; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which inc
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Imaging System Grid Overlay Display Error

    A software anomaly in the Boston Scientific AVVIGO+ Multi-Modality Guidance System causes an incorrect grid overlay display when used with certain catheters. The wrong overlay may lead to incorrect measurement references during imaging procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO+ BRIDGE TO INTEGRATED KIT; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    AVVIGO+ Medical Imaging System Grid Overlay Display Software Error

    Boston Scientific is recalling AVVIGO+ Multi-Modality Guidance Systems due to a software anomaly that displays a 9-grid overlay instead of the correct 15-grid overlay during Live and Record modes. This incorrect display could affect imaging measurements during procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST EU + RO; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalitie
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Multi-Modality Guidance System Grid Display Software Defect

    Boston Scientific is recalling the AVVIGO+ Guidance System for a software defect that displays incorrect grid overlays when connected to OptiCross 18 catheters. The error affects Live and Record modes only; no injuries have been reported.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL - ZERO COST US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalitie
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Cardiac Imaging System Grid Display Error

    A software anomaly in Boston Scientific's AVVIGO+ system causes an incorrect 9-grid overlay to display instead of the correct 15-grid overlay when connected to OptiCross 18 catheters, potentially affecting clinical interpretation of vascular imaging.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE EU ROW; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Imaging System Grid Display Error

    Boston Scientific is recalling AVVIGO+ Multi-Modality Guidance Systems due to a software error causing incorrect grid overlay display when used with OptiCross 18 catheters, which could affect measurement accuracy.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includ
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific imaging system software error in grid overlay display

    Boston Scientific AVVIGO+ imaging system has a software error causing incorrect grid overlay display in Live and Record modes, potentially affecting depth reference during vascular procedures. 2 units are recalled.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includ
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific Medical Device Software Error Affects Vessel Imaging Display

    The Boston Scientific AVVIGO+ Guidance System displays an incorrect grid overlay when used with the OptiCross 18 catheter in Live or Record mode, potentially affecting depth reference settings. No injuries have been reported.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM - ZERO COST JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ guidance system recalled for grid overlay display error

    Boston Scientific's AVVIGO+ guidance system is recalled because a software glitch causes an incorrect grid overlay display when used with an OptiCross 18 catheter. The correct overlay appears in Review mode, and no patient injuries have been reported.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Guidance System Recalled for Grid Overlay Display Error

    Boston Scientific is recalling AVVIGO+ systems due to a software anomaly that displays an incorrect grid overlay when connected to OptiCross 18 catheters, potentially leading to vessel imaging misinterpretation.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EU + ROW; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which inc
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Medical Device Grid Overlay Display Error

    Boston Scientific recalls 161 AVVIGO+ Multi-Modality Guidance Systems due to a software error causing an incorrect grid overlay display in Live and Record modes. The error affects measurement reference accuracy during intravascular imaging procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    AVVIGO+ Multi-Modality Guidance System software defect causes incorrect grid display

    Boston Scientific recalled the AVVIGO+ Multi-Modality Guidance System because a software defect causes incorrect grid display in Live and Record modes, potentially affecting vessel imaging guidance for physicians using OptiCross 18 catheters.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM - ZERO COST US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities whi
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Medical imaging system software defect causes incorrect grid overlay display

    Boston Scientific recalled the AVVIGO+ imaging guidance system due to a software defect that displays a 9-grid overlay instead of the correct 15-grid overlay when connected to OptiCross 18 catheters in Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Imaging System Grid Display Overlay Error

    Boston Scientific is recalling 5 units of its AVVIGO+ Guidance System due to a software error that displays an incorrect grid overlay when connected to OptiCross 18 Peripheral Imaging Catheters in Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO + INT Z UPGRADE JP; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ System Grid Overlay Display Software Error

    Boston Scientific is recalling the AVVIGO+ Multi-Modality Guidance System due to a software error that displays an incorrect 9-grid overlay instead of the correct 15-grid overlay in Live or Record mode, affecting measurement accuracy with OptiCross 18 catheters.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO + MOB Z UPGRADE JP; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Guidance System Grid Mark Overlay Defect

    Boston Scientific is recalling the AVVIGO+ Multi-Modality Guidance System due to a software defect that displays an incorrect 9-grid mark overlay instead of the correct 15-grid mark overlay when used with OptiCross 18 Peripheral Imaging Catheters in Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which incl
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Multi-Modality Imaging System Grid Overlay Display Error

    Boston Scientific is recalling AVVIGO+ Multi-Modality Guidance Systems due to a software error that displays incorrect 9-grid marks instead of 15-grid marks in Live and Record modes. This display error could affect physicians' ability to set proper reference depths during intravascular imaging procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ System Grid Overlay Display Software Error

    Boston Scientific AVVIGO+ Multi-Modality Guidance System displays incorrect grid overlay when connected to OptiCross 18 catheters in Live or Record mode due to a software anomaly. The wrong overlay could affect reference depth settings used in vascular imaging procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL EU + ROW; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities whic
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-10-23

    Boston Scientific AVVIGO+ Guidance System Grid Mark Display Error

    Boston Scientific is recalling the AVVIGO+ Guidance System due to a software bug displaying incorrect grid marks during live imaging. The device shows 9 grid marks instead of 15, which could affect accurate vessel measurement in cardiac interventions.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities whi
    Category
    Distribution
    1 state