FDA (Devices) recall action 95289

Philips North America Llc recalled 13 products on 2024-10-02

The FDA (Devices) publishes a recall like this one as 13 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 13 pages.

Products
13
Announced
2024-10-02
Critical
0
Units
385

Why these products were recalled

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–13 of 13

  • HighFDA (Devices)··2024-10-02

    MRI Scanner Loose Screws May Block Tabletop and Delay Diagnosis

    The Ingenia Elition X MRI device may have loose screws that interfere with the horizontal tabletop, potentially delaying patient diagnosis and causing anxiety during extended bore exposure.

    Product
    Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782151; 2) 782119; 3) 781358; 4) 782107; 5)…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    Philips Ingenia 1.5T MRI Device Recall: Loose Screws May Block Tabletop

    Philips recalls certain Ingenia 1.5T MRI devices due to loose assembly screws that may block the patient tabletop, potentially delaying diagnosis and causing patient anxiety.

    Product
    Ingenia 1.5T -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782115 2) 782101 3) 782140
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI Diagnostic Device Recalls Due to Loose Screw Assembly Risk

    Philips recalls Evolution Upgrade 3.0T MRI devices due to loose screws that may block the tabletop and delay diagnosis. Devices with the affected serial numbers should be taken out of service.

    Product
    Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    Philips MR 7700 MRI Scanner Recall for Loose Screws

    Philips is recalling 7 MR 7700 MRI scanners because loose screws may block the tabletop, potentially delaying diagnosis or causing patient anxiety.

    Product
    MR 7700 -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782120
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI Diagnostic Device Screws May Loosen, Block Tabletop

    Philips SmartPath to dStream 1.5T MRI devices may have loose assembly screws that could block the diagnostic tabletop and delay imaging. 32 units distributed worldwide.

    Product
    SmartPath to dStream for 1.5T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782112; 2) 781260
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2024-10-02

    Philips Ingenia 3.0T CX MRI Machine Carrier Assembly Screws May Loosen

    Philips Ingenia 3.0T CX MRI machines may have loose screws in the RF carrier assembly that could protrude and interfere with the tabletop mechanism, potentially delaying patient scans.

    Product
    Ingenia 3.0T CX - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782105
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI Diagnostic Device Recalled for Loose Assembly Screws Blocking Equipment

    Philips Ingenia Elition S MRI machines (10 units) are recalled because assembly screws may come loose and interfere with the horizontal tabletop, potentially delaying diagnosis and causing patient anxiety.

    Product
    Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782150 2) 782106
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    Philips Ingenia 3.0T MRI Scanner Tabletop Blockage Risk

    Philips is recalling 33 Ingenia 3.0T MRI scanners whose Integrated Radio Frequency Carrier assembly screws may come loose and block the horizontal tabletop, potentially causing diagnostic delays.

    Product
    Ingenia 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782103
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    Philips Ingenia Ambition X MRI Scanner Recalled Due to Loose Screws

    Philips is recalling 63 units of the Ingenia Ambition X MRI scanner due to loose screws in the IRF carrier assembly that may block the imaging tabletop and potentially delay diagnosis.

    Product
    Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782109
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI Diagnostic Device Loose Screws May Block Patient Tabletop

    Philips SmartPath dStream MRI devices have loose screws in the RF carrier assembly that may interfere with the moving tabletop, potentially delaying diagnosis. Affected units should be inspected and repaired immediately.

    Product
    SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI diagnostic device screws may loosen and block tabletop

    Philips is recalling the Ingenia Elition X MRI device due to loose screws in the RF carrier assembly that could block the tabletop and delay diagnosis or cause patient anxiety.

    Product
    SmartPath to Ingenia Elition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782118
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    MRI Device Ingenia Ambition S Screws May Loosen and Block Tabletop

    Assembly screws in Philips Ingenia Ambition S MRI systems may loosen and protrude, potentially blocking the patient tabletop and delaying diagnosis. Philips is recalling 64 affected units worldwide.

    Product
    Ingenia Ambition S - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782108 2) 781359 3) 782133
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    Philips MR 5300 MRI System Recalled for Loose Assembly Screws

    Philips is recalling 68 MR 5300 MRI diagnostic systems worldwide because loose screws in the RF carrier assembly could block equipment movement and delay patient diagnosis.