The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-10-02
Critical
0
Units
9,333

Why these products were recalled

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2024-10-02

    Siemens Diagnostic Ultrasound Systems Recalled for Incorrect Unit Conversion

    Siemens ACUSON Maple ultrasound systems may perform incorrect unit conversion in software, producing values 10 times smaller than actual. This could lead to misdiagnosis or inappropriate patient management.

    Product
    ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
    Category
    Distribution
    42 states
  • HighFDA (Devices)··2024-10-02

    ACUSON Redwood Ultrasound Systems Recall for Software Unit Conversion Error

    Siemens ACUSON Redwood ultrasound systems may display measurements 10 times smaller than actual if operators change the software from factory defaults to milliliters per second or minute. This could lead to misdiagnosis and poor patient management.

    Product
    ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software
    Category
    Distribution
    42 states
  • HighFDA (Devices)··2024-10-02

    Siemens ACUSON Juniper ultrasound systems software unit conversion error

    Siemens is recalling ACUSON Juniper ultrasound systems that may perform incorrect unit conversions when configured with non-default settings, potentially producing measurements 10 times smaller than actual values and risking patient misdiagnosis.

    Product
    ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
    Category