The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-09-25
Critical
0
Units
38,128

Why these products were recalled

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2024-09-25

    Diagnostic test reagent may produce inaccurately low patient results

    Siemens Healthcare's ADVIA Chemistry Microalbumin diagnostic test may produce falsely low results. Eight lots do not meet High-Dose Hook Effect and Prozone Effect specifications, potentially affecting microalbumin detection accuracy.

    Product
    ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-09-25

    Siemens Atellica Microalbumin Diagnostic Test Kits Recalled for Measurement Error

    Siemens Healthcare Diagnostics is recalling Atellica CH and ADVIA Chemistry Microalbumin test kits worldwide because they may not properly measure microalbumin in urine samples, potentially causing falsely low results that could delay diagnosis of kidney disease.