FDA (Devices) recall action 95216

Wipro GE Healthcare Private Ltd. recalled 2 products on 2024-10-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-10-02
Critical
0
Units
138

Why these products were recalled

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2024-10-02

    GE Giraffe Warmer Recalled for Incomplete Electrical Safety Testing

    Wipro GE Healthcare is recalling specific GE Giraffe Warmers because incomplete electrical safety testing during manufacturing may allow leakage current to exceed safety limits.

    Product
    GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-02

    GE Panda iRES infant warmer electrical safety testing incomplete

    GE Panda iRES infant warmers may have incomplete electrical safety testing that could cause leakage current to exceed safe limits. The company is recalling 137 units distributed in the US and internationally.