The Recall Desk

FDA (Devices) recall action 95127

MEDLINE INDUSTRIES, LP - Northfield recalled 4 products on 2024-09-18

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-09-18
Critical
0
Units
1,093

Why these products were recalled

A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2024-09-18

    Anesthesia Turnover Kit with Non-Sterile Mastisol Component Recalled

    Medline Industries recalls Anesthesia Turnover Kit Pack Number DYNJAA0377 due to a non-sterile Mastisol component placed in the sterile kit without proper labeling. This poses an infection risk if users assume the component is sterile.

    Product
    ANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-09-18

    Sterile Anesthesia Kits Recalled for Non-Sterile Component Mixing

    Medline Industries recalled anesthesia kits containing an unlabeled non-sterile Mastisol component that could be mistakenly assumed sterile during use, creating infection risk. The recall affects 315 units distributed in Texas, Illinois, and California.

    Product
    ANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377A
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-09-18

    Sterile Surgical Kit Contains Non-Sterile Component Without Label

    A non-sterile Mastisol component was included in certain sterile craniotomy packs without proper labeling. If users assume the component is sterile, it could increase infection risk during surgery.

    Product
    CRANIOTOMY PACK, Pack Number DYNJ46552O
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-09-18

    Medline Sterile Medical Kit with Mislabeled Non-sterile Component Recalled

    Medline's KIT PEDI CRANI LAMI sterile kits contain a non-sterile Mastisol component not labeled as such, raising infection risk. 485 units were distributed in Texas, Illinois, and California.

    Product
    KIT PEDI CRANI LAMI, Pack Number DYNJ907372A
    Category
    Distribution
    3 states