FDA (Devices) recall action 95121

Medtronic MiniMed, Inc. recalled 3 products on 2024-10-09

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-10-09
Critical
0
Units
785,513

Why these products were recalled

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stoppi…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • SevereFDA (Devices)··2024-10-09

    MiniMed Insulin Pumps Recalled for Battery Failure After Physical Impact

    MiniMed insulin pumps are recalled because physical impact can damage internal components, reduce battery life, and cause unexpected device shutdown that interrupts insulin delivery.

    Product
    MiniMed 620G Insulin Pump, REF: MMT-1510, MMT-1710, MMT-1750; MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 640G Insulin Pump, REF: MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, MMT-1752; MiniMed 720G Insulin Pump, REF…
    Category
    Distribution
    53 states
  • SevereFDA (Devices)··2024-10-09

    MiniMed Insulin Pumps May Stop Delivering Insulin After Physical Impact

    Medtronic MiniMed insulin pumps (670G, 770G, 780G) that have been dropped or bumped may develop internal battery defects causing premature failure and interrupted insulin delivery, risking dangerously high blood sugar.

    Product
    MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884…
    Category
    Distribution
    53 states
  • SevereFDA (Devices)··2024-10-09

    MiniMed Insulin Pumps Recalled for Battery Depletion and Delivery Risk

    Medtronic MiniMed 630G and 700G insulin pumps may experience battery depletion and device shutdown after physical impact, interrupting insulin delivery in diabetic patients.

    Product
    MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851
    Category
    Distribution
    53 states