Beckman Coulter Biomedical GmbH recalled 3 products on 2024-09-11

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-09-11
Critical
0
Units

Why these products were recalled

The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • HighFDA (Devices)··2024-09-11

    Beckman Power Express Centrifuge Sample Holder May Detach During Operation

    An adhesive joint failure in the PE Centrifuge Temperature Controlled component can cause the metal part to detach inside the centrifuge, risking sample destruction and potential exposure of laboratory technicians to biohazardous material.

    Product
    PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-09-11

    Medical Device Sample Holder Recalled for Adhesive Joint Failure

    Beckman Coulter Power Express sample holders can fail at the adhesive joint between plastic socket and metal component. This may cause the metal part to detach inside the centrifuge, potentially damaging equipment and exposing lab staff to biohazardous material.

    Product
    Power Express, REF B90918
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-09-11

    Power Express PE Centrifuge adhesive joint failure may cause component detachment

    The adhesive joint in this medical device component can fail, causing the metal part to detach. This may damage the centrifuge, destroy patient samples, and potentially expose lab staff to biohazardous material.

    Product
    PE Centrifuge, REF: B36365, a component of the Power Express
    Category
    Distribution
    0 states