The Recall Desk

FDA (Devices) recall action 95019

BioFire Diagnostics, LLC recalled 2 products on 2024-09-18

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-09-18
Critical
0
Units
4,440

Why these products were recalled

Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-09-18

    BioFire FILMARRAY TORCH Module Recall Due to Calibration Errors

    BioFire is recalling FILMARRAY TORCH modules because calibration errors may cause false positive or negative results for pathogens like Salmonella and MRSA.

    Product
    FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2024-09-18

    FilmArray 2.0 Diagnostic Instruments Recalled for Calibration Error Affecting Test Results

    BioFire Diagnostics is recalling FilmArray 2.0 diagnostic instruments that may have been incorrectly calibrated, leading to false positive and false negative results for infectious disease tests including Salmonella, Giardia, Norovirus, and MRSA.

    Product
    FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
    Category
    Distribution
    52 states