FDA (Devices) recall action 94979

DKK Dai-Ichi Shomei Co., Ltd. recalled 4 products on 2024-08-21

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-08-21
Critical
0
Units
2,208

Why these products were recalled

The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2024-08-21

    Aurora 7 Surgical Light Heads Recalled for Cracked Weld Frames

    DKK Dai-Ichi Shomei is recalling 555 Aurora 7 surgical light heads because frame welds can crack, causing the frame to sag, the housing to crack, or the light head to detach.

    Product
    Aurora Series, Aurora 7 lighthead; Models: LED7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-21

    Surgical light heads may detach due to weld defects

    DKK Dai-Ichi Shomei is recalling 293 Aurora surgical light heads because the frame welds can crack, potentially causing the light head to detach during use. No injuries have been reported.

    Product
    Aurora 3 Series, Aurora 7 lighthead; Models: AUT7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-21

    Surgical Lights Recalled for Possible Weld Defect and Frame Detachment

    DKK Dai-Ichi Shomei Co., Ltd. is recalling 822 Aurora surgical light heads due to a weld defect that could cause the light head to sag, crack, or detach during use.

    Product
    Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-21

    Surgical Lighting Equipment Recalled for Cracked Welds and Frame Detachment Risk

    Aurora Series surgical light heads are recalled due to cracked welds on the frame. The defects can cause the frame to sag, housing to crack, or the light head to detach.

    Product
    Aurora 4 Series, Aurora 7 lighthead; Models: AUA7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Distribution
    0 states