The Recall Desk

FDA (Devices) recall action 94928

Hitachi America, Ltd., Particle Therapy Division recalled 3 products on 2024-09-18

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-09-18
Critical
0
Units
3

Why these products were recalled

During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2024-09-18

    Proton therapy system may display insufficient CBCT images during positioning

    Hitachi PROBEAT-FR proton therapy systems may display an insufficient number of CBCT images during 3D positioning when the Float Image Selection button is pressed, potentially affecting treatment planning accuracy.

    Product
    Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-09-18

    Proton Therapy System May Display Insufficient Images During Treatment Planning

    FDA recalls the PROBEAT-V proton beam therapy system due to a software issue where the 3D image positioning mode may display insufficient images, potentially affecting treatment planning accuracy.

    Product
    Brand Name: PROBEAT Product Name: PROBEAT-V Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2024-09-18

    Proton Beam Therapy System Displays Insufficient CBCT Images on Button Press

    Hitachi's PROBEAT-CR proton therapy system may display insufficient CBCT images during image positioning when the Float Image Selection button is pressed. This could affect treatment planning accuracy.

    Product
    Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam…
    Category
    Distribution
    3 states