The Recall Desk

FDA (Devices) recall action 94908

American Contract Systems, Inc. recalled 8 products on 2024-08-14

The FDA (Devices) publishes a recall like this one as 8 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 8 pages.

Products
8
Announced
2024-08-14
Critical
0
Units
596

Why these products were recalled

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–8 of 8

  • HighFDA (Devices)··2024-08-14

    Gyn Laparoscopy Kits Recalled Due to Inadequate Sterilization Process

    American Contract Systems is recalling 29 gynecological laparoscopy surgical kits due to insufficient aeration during sterilization, which may compromise sterility and increase infection risk for patients.

    Product
    American Contract Systems Gyn Laparoscopy convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    Surgical Convenience Kits Recalled for Potential Sterilization Failure

    American Contract Systems is recalling three surgical convenience kits due to potential sterilization failure from insufficient aeration time during processing.

    Product
    American Contract Systems Extremity convenience kit, American Contract Systems Hand Pack convenience kit, and American Contract Systems Total Knee Pack convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    Medical Procedure Kit Recalled Over Potential Sterilization Compromise

    American Contract Systems is recalling Special Procedure Pack convenience kits due to insufficient aeration time during sterilization, which may have compromised sterility. Distributed in California, Kansas, Maine, Minnesota, Missouri, Nebraska, and South Dakota.

    Product
    American Contract Systems Special Procedure Pack convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    American Contract Systems Hand Pack Convenience Kit Sterilization Failure

    American Contract Systems is recalling Hand Pack convenience kits that may not be sterile due to insufficient aeration time during sterilization. Affected devices distributed in CA, KS, ME, MN, MO, NE, and SD may pose a safety risk.

    Product
    American Contract Systems Hand Pack convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    American Contract Systems Spine Pack kits recalled for insufficient sterilization aeration

    American Contract Systems is recalling Spine Pack convenience kits (Lot 980241) due to incomplete post-sterilization aeration. The affected kits were distributed in California, Kansas, Maine, Minnesota, Missouri, Nebraska, and South Dakota.

    Product
    American Contract Systems Spine Pack convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    American Contract Systems surgical kits recalled for sterilization defect

    American Contract Systems Laparoscopy and Lap Chole surgical kits are being recalled because insufficient aeration time after sterilization may have compromised their sterility. The kits were distributed in seven U.S. states.

    Product
    American Contract Systems Laparoscopy Litho Std Pack convenience kit and American Contract Systems Lap Chole convenience kit
    Category
    Distribution
    7 states
  • HighFDA (Devices)··2024-08-14

    Biopsy Drape Pack sterilization failure recall from American Contract Systems

    American Contract Systems is recalling 224 Biopsy Drape Pack convenience kits due to insufficient aeration time during sterilization, which may have compromised the sterility of affected units. Lot 981241 distributed in CA, KS, ME, MN, MO, NE, and SD is affected.

    Product
    American Contract Systems Biopsy Drape Pack convenience kit
    Category
    Distribution
    7 states