The Recall Desk

FDA (Devices) recall action 94890

B Braun Medical Inc recalled 5 products on 2024-07-31

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2024-07-31
Critical
0
Units
32,004

Why these products were recalled

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–5 of 5

  • HighFDA (Devices)··2024-07-31

    Perfusor Space syringe pump recalled for Cardinal Health syringe incompatibility

    B Braun is recalling Perfusor Space syringe pumps because manufacturing changes to Cardinal Health Monoject syringes may cause incompatibility, resulting in dosing errors or delayed treatment.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each Component: N/A
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-31

    Perfusor Space Syringe Pump Recalled Due to Monoject Syringe Incompatibility

    B Braun Medical's Perfusor Space Syringe Pump may not properly recognize Cardinal Health Monoject syringes due to manufacturing changes, risking medication overdose, underdose, or therapy delays.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-31

    Perfusor Space Syringe Pump Recalled for Monoject Syringe Incompatibility Risk

    B Braun's Perfusor Space infusion pump may be incompatible with Cardinal Health Monoject syringes affected by manufacturing changes, creating risks of overdose, underdose, and delayed therapy.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-31

    Perfusor Space infusion pump syringe incompatibility may cause dosing error

    B Braun is recalling Perfusor Space infusion pump kits because Cardinal Health Monoject syringes have manufacturing changes that make them incompatible, potentially causing overdose, underdose, or delayed therapy.

    Product
    Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-31

    Infusion Pump Syringe Compatibility Issue May Cause Overdose or Underdose

    B Braun is recalling Perfusor Space infusion pumps due to syringe compatibility issues with Cardinal Health Monoject syringes. The incompatibility may cause overdose, underdose, therapy delays, and alarm failures.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A
    Category
    Distribution
    0 states