The Recall Desk

FDA (Devices) recall action 94874

GE Medical Systems, LLC recalled 26 products on 2024-08-14

The FDA (Devices) publishes a recall like this one as 26 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 26 pages.

Products
26
Announced
2024-08-14
Critical
0
Units
16,043

Why these products were recalled

GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 26

  • HighFDA (Devices)··2024-08-14

    MRI System Software Error Can Overheat Patient Implants

    GE Medical Systems recalled three SIGNA 7.0T MRI systems due to a software error that can exceed preset safety limits during imaging, potentially overheating implants in patients under certain conditions.

    Product
    SIGNA 7.0T, with affected software versions: 7T29.1, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA MR355 MRI System Recalled for Potential Implant Overheating Risk

    GE SIGNA MR355 MRI systems running software versions SV25.5 or SV25.6 may cause implant overheating in certain scanning conditions. Risk occurs when using specific imaging sequences with the Optimized T2 FLAIR feature disabled.

    Product
    SIGNA MR355, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Prime MR system can exceed RF safety limits in Low SAR mode

    GE Medical Systems is recalling 72 SIGNA Prime MR imaging systems worldwide. Under specific imaging conditions, the systems can exceed prescribed radiofrequency safety limits, potentially causing MR conditional implants to overheat.

    Product
    SIGNA Prime, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Premier MRI system software overheating risk to implants

    GE Medical Systems is recalling 404 SIGNA Premier MRI systems worldwide due to a software issue where radiofrequency energy can exceed prescribed safety limits during specific imaging sequences, potentially overheating MR-conditional implants.

    Product
    SIGNA Premier, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    MR System Software Defect May Cause Implant Overheating During Scanning

    GE SIGNA Explorer MR imaging systems with affected software versions may overheat implants during certain scan sequences. 2,528 units affected worldwide; no injuries reported.

    Product
    SIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Pioneer MR Systems May Overheat Patient Implants During Scanning

    GE SIGNA Pioneer MR systems with certain software versions may allow radiofrequency exposure to exceed prescribed limits in Low SAR Mode when using specific imaging sequences, potentially causing overheating of MR conditional implants.

    Product
    SIGNA Pioneer, with affected software versions: PX25.0 to PX25.5, PX26.0, PX26.1, PX28.0 to PX28.4, PX29.0, PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    SIGNA UHP MRI System Software Versions May Cause Implant Overheating

    GE Medical Systems is recalling SIGNA UHP MRI systems with software versions RX28.0 or MR30.1. A configuration issue can cause radiofrequency output to exceed prescribed limits, potentially overheating MR conditional implants.

    Product
    SIGNA UHP, with affected software versions: RX28.0, MR30.1; Nuclear Magnetic Resonance Imaging System - research use only
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    SIGNA MAGNUS MRI Software Can Overheat Implanted Devices

    GE Medical Systems recalled SIGNA MAGNUS MRI systems with software versions MR29.1 or RX29.1. Under specific imaging conditions in Low SAR Mode, the system may exceed safe radiofrequency power limits and cause implanted medical devices to overheat.

    Product
    SIGNA MAGNUS, with affected software versions: MR29.1, RX29.1; Nuclear Magnetic Resonance Imaging System - research use only
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Artist MR system may overheat implants in certain sequences

    GE Medical Systems is recalling SIGNA Artist MR imaging systems because they may produce radiofrequency power exceeding user-set limits when using certain imaging sequences in Low SAR Mode, potentially causing MR-conditional implants to overheat.

    Product
    SIGNA Artist, with affected software versions: DV26.0, DV27.1, DV28.1, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    SIGNA MR380 MRI system software may exceed implant heating limits

    GE SIGNA MR380 MRI systems with software versions SV25.5 or SV25.6 may exceed radiofrequency heating limits during certain imaging sequences, potentially causing MR-conditional implants to overheat. The issue occurs in Low SAR Mode when specific T2 FLAIR sequences are used with the Optimized T2 FLAIR option disabled.

    Product
    SIGNA MR380, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA HDxt 3.0T MRI systems may overheat implanted devices

    GE Medical Systems is recalling 3.0T SIGNA HDxt MRI systems because radiofrequency energy can exceed safety limits under specific operating conditions, potentially overheating implanted medical devices in patients.

    Product
    3.0T SIGNA HDxt Family, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1HD16.0_V03, HD23.0_V03; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA MR360 MRI System Software Defect May Overheat Implants

    GE SIGNA MR360 MRI systems with software versions SV25.5 or SV25.6 may exceed prescribed RF exposure limits in certain scan configurations, potentially overheating MR-conditional implants. The recall affects 81 units worldwide.

    Product
    SIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    MRI System Software Error Can Cause Implant Overheating in Low SAR Mode

    GE HealthCare's Discovery MR950 MRI system may incorrectly calculate heating safety limits when using certain imaging sequences in Low SAR Mode, potentially causing overheating of MR-conditional implants.

    Product
    Discovery MR950, with affected software versions: 7T23.0; Nuclear Magnetic Resonance Imaging System - research use only
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE Optima MR360 MRI System Software May Overheat Implants

    GE Optima MR360 MRI systems with affected software versions may exceed safe radiofrequency energy limits, potentially overheating MR-conditional implants when specific imaging sequences are used. 469 units are affected worldwide.

    Product
    Optima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE Discovery MR750 MRI System Software May Cause Implant Overheating

    GE HealthCare is recalling certain MR750 MRI systems with specific software versions. When using particular imaging sequences in Low SAR Mode without the Optimized T2 FLAIR option, the radiofrequency field can exceed safe limits, risking overheating of MR-conditional implants.

    Product
    Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Voyager MRI systems risk implant overheating under specific configurations

    GE SIGNA Voyager MRI systems may exceed prescribed radiofrequency limits under specific operating conditions, risking overheating of MR conditional implants. The issue occurs with certain imaging sequences when the Optimized T2 FLAIR option is disabled.

    Product
    SIGNA Voyager, with affected software versions: PX26.0 to PX26.6, VX28.0, VX29.1, VX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    SIGNA PET/MR diagnostic imaging system poses implant overheating risk

    GE HealthCare recalled SIGNA PET/MR systems with specific software versions due to a configuration issue that can cause RF energy to exceed safety limits, potentially overheating MR-conditional implants.

    Product
    SIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA 1.5T MR Systems May Overheat Implants in Low SAR Mode

    GE Medical Systems recalled 2,146 SIGNA 1.5T MR systems that may overheat implants during certain imaging sequences in Low SAR mode due to a software defect in radiofrequency safety calculations.

    Product
    1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE SIGNA Creator MRI systems software defect risks patient implant overheating

    GE HealthCare recalled certain SIGNA Creator MRI systems due to a software defect where the predicted magnetic field strength can exceed prescribed limits in Low SAR Mode with specific imaging sequences, potentially causing overheating of MR-conditional implants.

    Product
    SIGNA Creator, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE MR450w 1.5T MRI software defect may cause implant overheating

    GE MR450w 1.5T MRI systems may overheat MR-conditional implants when using specific imaging sequences with Low SAR Mode. No injuries reported; 1,607 units affected worldwide.

    Product
    Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    MRI System Software Miscalculation May Overheat Patient Implants

    A software error in GE SIGNA Hero MRI systems can cause overheating of MR conditional implants under specific conditions. The recall affects 95 units distributed worldwide.

    Product
    SIGNA Hero, with affected software versions: PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    Discovery MR450 MRI System Can Overheat Implants Under Specific Conditions

    GE Discovery MR450 MRI systems with certain software versions can cause implant overheating under specific scanning conditions. 23 units worldwide require firmware updates.

    Product
    Discovery MR450 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    MRI System Software Issue Can Overheat Implanted Medical Devices

    GE HealthCare recalled 23 SIGNA Victor MRI systems due to a software issue that can cause MR conditional implants to overheat. The defect occurs under specific imaging mode and configuration combinations.

    Product
    SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    GE Discovery MR750w MRI System Software May Overheat Implants

    GE HealthCare is recalling certain Discovery MR750w 3.0T MRI systems where the magnetic field strength (B1+RMS) can exceed the prescribed limit during specific imaging sequences, potentially causing overheating of MR-conditional implants.

    Product
    Discovery MR750w 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.5, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-08-14

    SIGNA Architect MRI System Software Can Exceed Implant Heating Limits

    GE Medical Systems is recalling SIGNA Architect MRI systems in which software may allow B1+ radiofrequency exposure to exceed user-prescribed limits during certain T2 FLAIR imaging sequences, risking MR-conditional implant overheating.

    Product
    SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states