Merit Medical Systems, Inc. recalled 23 products on 2024-08-21

The FDA (Devices) publishes a recall like this one as 23 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 23 pages.

Products
23
Announced
2024-08-21
Critical
0
Units
2,890,251

Why these products were recalled

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–23 of 23

  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits Recalled for Syringe Leaks and Breakage

    Merit Medical is recalling medical convenience kits with defective syringes due to identified leaks, breakage, and quality issues that may pose a risk to patient health during procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K05-02166B K05T-00380C K05T-01674B K05T-01863 K05T-01941C K05T-01955 K05T-02138 K05T-02159 K05T-02272G…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical catheter extractor kits recalled for syringe quality defects

    Merit Medical is recalling 3,631 catheter extractor convenience kits due to defective syringes with leaks and breakage. The identified quality issues may compromise patient safety during catheter removal.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Outake, REF CET200 and CET300, The Catheter Extractor is a catheter accessory device intended for minimally invasive removal of tunneled long term catheters.…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits with Recalled Syringes Pose Quality Risks

    Merit Medical is recalling convenience kits containing syringes with quality defects including leaks and breakage that may pose patient safety risks in cardiac and vascular procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/angioplasty kit), REF: K12T-10773 K12T-09284D K12T-10621 K12T-10800 K12T-12491 To support various vascular or cardiac…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical pericardiocentesis kits recalled for syringe defects

    Merit Medical convenience kits containing plastic syringes have been recalled nationwide. The syringes have quality issues including leaks and breakage that may pose a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: PC (Pericardiocentesis Tray - Fluid drainage tray), REF: PC101/B PC801 PC801/A PC802 PC802/A The Pericardiocentesis Kit is intended to withdraw fluid from…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical convenience kits recalled for syringe leaks and breakage

    Merit Medical is recalling 1,390 convenience kits containing plastic syringes with identified quality issues including leaks and breakage. The kits, distributed nationwide, may pose a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular administration set is a device used to administer fluids from a container to a patient's vascular…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits with Defective Syringes Recalled

    Merit Medical convenience kits containing plastic syringes are recalled due to leaks, breakage, and quality issues that may pose risks to patients undergoing cardiac and vascular procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF: K05-00203M K05-00511J K05-00547J K05-00612Q K05-00819A K05-01030H K05-01274D K05-01646 K05-01688A K05-01717B K05-01880K…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Syringes in Convenience Kits Recalled for Leaks and Breakage

    Merit Medical convenience kits containing plastic syringes are recalled due to leaks, breakage, and quality defects that may pose a risk to patient health. Approximately 3,647 kits were distributed nationwide.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K02 (Custom Syringe Kit), REF: K02-00873A, Inject fluids into, or withdraw fluids from, the body.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical convenience kits with syringe quality issues recalled nationwide

    Merit Medical is recalling convenience kits containing syringes with quality issues including leaks and breakage that may pose a risk to patients receiving intravenous fluid administration.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K08-PI (Custom Fluid Management Kit), REF: K08-02042 K08-02948 K08-03026 K08-03119 An intravascular administration set is a device used to administer…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical convenience kits recalled for syringe leaks and breakage

    Merit Medical convenience kits containing recalled syringes have quality issues including leaks and breakage that may pose a risk to patient health. The recall affects 2,388 kits distributed nationwide.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Rapid Response Waste Management System Kit, REF BRC-BAL-INT-SL/B An endoscope and accessories is a device used to provide access, illumination, and allow…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Pressure Monitoring Kits with Defective Syringes Recalled

    Merit Medical is recalling 69 convenience kits containing defective syringes that may leak or break, potentially risking patient safety during medical procedures. The kits were distributed nationwide.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K11 (Custom Pressure Monitoring Kit), REF: K11-00382C To support fluid administration into the body and physiological pressure monitoring. To support various…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Catheter Kits Recalled for Syringe Leaks and Breakage Issues

    Merit Medical convenience kits containing defective plastic syringes are being recalled due to leaks, breakage, and quality issues that may harm patients. The recall affects approximately 4,027 kits distributed nationwide.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG/Worley advanced CSG (Introducer Catheter kit), REF: FCL-083-02/A WOR-CSG-B1-09/A WOR-CSG-B2-09/A WOR-CSG-BL1-09/A WOR-CSG-E-90/A For the…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits Recalled for Defective Plastic Syringes

    Merit Medical is recalling 7,442 convenience kits containing defective syringes that may leak or break, posing a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: ReSolve (Catheter and Tip Suction kit), REF: RTT14038MB/A The ReSolve Thoracostomy Tray is indicated for the percutaneous removal of air and fluid from the…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Cardiac Catheter Kits Recalled for Syringe Leaks and Breakage

    Merit Medical's Prelude SNAP-H cardiac catheter kits are being recalled because their syringes may leak or break during use, potentially affecting cardiac procedures. No illnesses have been reported.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP-H (Introducer Catheter kit), REF: PLSH-1006/A PLSH-1007/A PLSH-1008/A PLSH-1009.5/A PLSH-2506/A PLSH-2507/A For the introduction of various…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Convenience kits by Merit Medical recalled for syringe defects

    Merit Medical convenience kits containing defective syringes have been recalled nationwide. Identified quality issues including leaks and breakage may pose risks during cardiac and vascular procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K12T/12PI (Custom Pressure Monitoring Kit), REF: K12-03846 K12-04185A K12-03447 K12M-02489 K12-10056 K12-02489A K12-02097B K12-12335 K12-12192P…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    FDA recalls Merit Medical convenience kits due to defective syringes

    Merit Medical convenience kits containing plastic syringes are recalled due to identified leaks, breakage, and quality issues that may pose a risk to patient health. The affected kits are used for vascular and cardiac procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K09/K09T (Custom Manifold Kit), REF: K09-00526L K09-00807AH K09-01467AD K09-01467AF K09-01514U K09-01734L K09-02040C K09-02683F K09-02804B K09-03844F…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits With Recalled Syringes Pose Quality Defect Risks

    Merit Medical convenience kits containing recalled syringes are subject to a nationwide recall due to quality defects including leaks and breakage that may pose risks to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K10/K10-PI (Custom Waste Management Kit), REF: K10T-05815 K10T-05816 K10-04722C K10-03591KP K10-05530 K10-04291AP K10-04386CP K10T-05227 K10-05120P…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Medical paracentesis procedure kits recalled for syringe leaks and breakage

    Merit Medical is recalling paracentesis procedure kits nationwide containing syringes with identified leaks and breakage. The defects may pose a risk to patient health during medical procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Safety Paracentesis Procedure Tray (Paracentesis Tray kit), REF: SPPT-100/D SPPT-5F-10/D SPPT-5F-10L/B SPPT-5F-10S SPPT-5F-12PL SPPT-5F-17PL SPPT-5F-7/D…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits Recalled Due to Defective Syringes

    Merit Medical convenience kits containing recalled plastic syringes have quality issues including leaks and breakage that may pose risks during vascular and cardiac procedures. Approximately 5,700 affected kits were distributed nationwide.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/Angioplasty kit), REF: OPT0242-00 S1279-01 S2953 S2960 S6002 To support various vascular or cardiac diagnostic and…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits with Recalled Syringes Pose Leak Risk

    Merit Medical is recalling 900 convenience kits containing plastic syringes due to identified leaks, breakage, and quality issues that may pose a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG (Introducer Catheter), REF: FCL-069-00/A FCL-069-02/A FCL-069-03/A For the introduction of various types of pacing or defibrillator leads and…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical Convenience Kits with Defective Syringes Recalled

    Merit Medical convenience kits containing defective syringes are being recalled for leaks and breakage that may pose a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude Prestige (Introducer Catheter), REF: PLP-2505/A PLP-2514/A PLP-2516/A The introduction of various types of pacing/defibrillator leads and catheters…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical cardiac kits contain plastic syringes with quality defects

    Merit Medical convenience kits used in cardiac procedures contain plastic syringes with potential leaks and breakage. The defects were identified following an FDA Safety Alert issued March 19, 2024.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP (Introducer Catheter kit), REF: For the introduction of various types of pacing leads and catheters to the heart and coronary venous system.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical convenience kits recalled for syringe leaks and breakage risks

    Merit Medical is recalling convenience kits containing syringes with quality defects including leaks and breakage. The recalled kits may pose a risk to patients undergoing vascular and cardiac procedures.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K22C (Vein Tray Angiography kit), REF: K09T-11796F K12T-01558G K12T-01558H K12T-01822 K12T-01852L K12T-02006J K12T-02054 K12T-02091G K12T-02102B…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2024-08-21

    Merit Medical convenience kits with recalled plastic syringes may leak or break

    Convenience kits from Merit Medical Systems containing recalled plastic syringes have identified quality issues including leaks and breakage that may pose a risk to patient health.

    Product
    Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K09T-01138 K09T-04484A K09T-04757F K09T-04835H K09T-07256D K09T-07588B K09T-07611K K09T-07649K…
    Category
    Distribution
    51 states