FDA (Devices) recall action 94772

GE Vingmed Ultrasound As recalled 2 products on 2024-07-17

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-07-17
Critical
0
Units
8

Why these products were recalled

GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-07-17

    GE Vivid Ultrasound Scanner Displays Incorrect Patient Data

    GE Vivid ultrasound systems may display incorrect patient names and demographics due to slow DICOM response times, risking patient misidentification during imaging.

    Product
    GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW
    Category
    Distribution
    5 states
  • HighFDA (Devices)··2024-07-17

    GE Vivid Ultrasound Systems Display Incorrect Patient Names

    GE Vivid ultrasound systems may display incorrect patient information due to slow DICOM response times. Seven affected units were distributed to healthcare facilities in IL, MA, NM, NY, and OK.

    Product
    GE Vivid Ultrasound products, Model Numbers: a) Vivid E90, H45581LB; b) Vivid E95, H45581DA
    Category
    Distribution
    5 states