FDA (Devices) recall action 94647

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalled 3 products on 2024-06-05

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-06-05
Critical
0
Units
5

Why these products were recalled

When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2024-06-05

    Philips Allura Xper FD20 patient table brake may fail to engage after software update

    Philips Allura Xper FD20 X-ray systems with ADN7NT patient tables may have a pivot brake failure after software upgrade 8.1.100 is installed. This could cause unexpected table movements during use.

    Product
    Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-06-05

    Philips Allura Xper FD10 system pivot brake failure after software update

    Philips is recalling Allura Xper FD10 medical imaging systems with software version 8.1.100 because the pivot brake may fail to engage, potentially causing unexpected table movements during procedures.

    Product
    Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-06-05

    Philips Allura Xper FD20 Software Update Disables Pivot Brake on Patient Table

    A software upgrade to Philips Allura Xper FD20 Biplane systems causes the patient table pivot brake to fail to engage, potentially causing unexpected table movements during procedures.

    Product
    Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.