The Recall Desk

FDA (Devices) recall action 94510

Lucid Diagnostics, Inc. recalled 2 products on 2024-06-05

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-06-05
Critical
0
Units
377

Why these products were recalled

The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-06-05

    EsophaCap Sponge May Detach From String During Removal

    Lucid Diagnostics is recalling EsophaCap medical devices (220 units in MN and MD) because the sponge may detach from the string during removal, particularly at esophageal sphincters.

    Product
    Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: N/A
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2024-06-05

    EsophaCap esophageal diagnostic sponge recalled for detachment risk

    Lucid Diagnostics recalls EsophaCap esophageal diagnostic sponges due to risk of sponge detachment from string during device removal. No injuries have been reported.

    Product
    Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T20VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: No
    Category
    Distribution
    2 states