The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-05-22
Critical
0
Units
51

Why these products were recalled

Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-05-22

    Philips Ingenia 3.0T MR Systems May Not Warn of Low Ventilation Settings

    Philips Ingenia 3.0T MR systems with software R5.7.1 may fail to warn about low ventilation, potentially causing patient overheating or burning sensations. 19 units affected worldwide.

    Product
    Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117
    Category
    Distribution
    4 states
  • HighFDA (Devices)··2024-05-22

    MRI System Software May Fail to Warn of Inadequate Ventilation

    Philips Ingenia 1.5T MRI systems with software R5.7.1 may not provide adequate warning for low ventilation conditions, potentially causing increased patient body temperature or burning sensation.

    Product
    Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
    Category
    Distribution
    4 states