The Recall Desk

FDA (Devices) recall action 94437

Stryker Neurovascular recalled 8 products on 2024-06-05

The FDA (Devices) publishes a recall like this one as 8 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 8 pages.

Products
8
Announced
2024-06-05
Critical
0
Units
8,011

Why these products were recalled

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–8 of 8

  • HighFDA (Devices)··2024-06-05

    FlowGate2 neurovascular devices may exceed endotoxin limits due to inadequate testing

    Stryker's FlowGate2 neurovascular devices may have excessive endotoxin levels because manufacturing failed to meet required testing standards. High endotoxin can cause fever, shock, and aseptic meningitis.

    Product
    FlowGate2 8F x 85cm, Catalog: 90485 FlowGate2 8F x 95cm, Catalog: 90495
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Neurovascular devices recalled due to inadequate endotoxin testing

    Stryker Neurovascular is recalling 81 neurovascular devices due to inadequate endotoxin testing. The devices may contain excessive endotoxin levels, which can cause fever, shock, and aseptic meningitis.

    Product
    DAC 044 115cm, Catalog: 90760
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Stryker neurovascular devices recalled for inadequate endotoxin testing

    Stryker Neurovascular is recalling Trevo ProVue devices due to manufacturing testing deficiency that may result in unsafe endotoxin levels and risk of fever, shock, or aseptic meningitis.

    Product
    Trevo ProVue, Catalog: 90184
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Stryker Neurovascular Microcatheter Recall Due to Inadequate Endotoxin Testing

    Stryker Neurovascular is recalling the Trevo Trak 21 Microcatheter due to potential endotoxin contamination from inadequate testing. High endotoxin levels may increase risk of fever, shock, and aseptic meningitis.

    Product
    Trevo Trak 21 Microcatheter, Catalog: 90338
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Synchro Guide Wires Recalled for Inadequate Endotoxin Testing

    Stryker Neurovascular is recalling 2,859 Synchro guide wires in multiple configurations due to inadequate endotoxin testing during manufacturing. The devices may exceed acceptable endotoxin limits, increasing the risk of fever, shock, and aseptic meningitis.

    Product
    Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Trevo NXT neurovascular devices recalled for inadequate endotoxin testing

    Stryker Neurovascular is recalling Trevo NXT neurovascular intervention devices because they did not receive adequate endotoxin testing during manufacturing. High endotoxin levels may cause fever, shock, or aseptic meningitis.

    Product
    Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK - CE, Catalog: 91414 TREVO NXT 6X37+TRAK21 2-PACK - CE, Catalog: 91415 TREVO NXT 3X32+PRO14 2
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Trevo XP ProVue neurovascular retriever devices recalled for inadequate endotoxin testing

    Stryker Neurovascular's Trevo XP ProVue Retriever devices did not undergo adequate endotoxin testing and may exceed safe limits, potentially causing fever, shock, and aseptic meningitis.

    Product
    Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x 25, Catalog: 90186 Kits: TREVO XP 3X20+PRO14 2-PACK, Catalog: 80051 TREVO XP 4X20+PRO18 2-PACK
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2024-06-05

    Trevo Pro Microcatheters Recalled for Potential Endotoxin Contamination

    Stryker Neurovascular is recalling Trevo Pro 14 and 18 Microchethers worldwide due to non-compliance with endotoxin testing standards. High endotoxin levels could cause fever, shock, and meningitis.

    Product
    Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
    Category
    Distribution
    50 states