The Recall Desk

FDA (Devices) recall action 94291

EBI, LLC recalled 4 products on 2024-05-01

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-05-01
Critical
0
Units

Why these products were recalled

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

The agency’s own wording, quoted from the enforcement report.

1–4 of 4

  • HighFDA (Devices)··2024-05-01

    Biomet OrthoPak Bone Growth Stimulator System Recalled Due to Lead Wire Damage

    The Biomet OrthoPak Bone Growth Stimulator System is being recalled due to potential lead wire sheath damage that could prevent therapeutic signal delivery and delay patient treatment nationwide.

    Product
    Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones. Part Number: 1067718
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-01

    Spine Fusion Stimulator Recalled for Damaged Lead Wire Sheath

    The Biomet SpinalPak Spine Fusion Stimulator System is recalled for potential lead wire sheath damage that may prevent delivery of the therapeutic treatment signal. Devices distributed nationwide since May 2023 are affected.

    Product
    Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-05-01

    Replacement Lead Wires Recalled for Risk of Treatment Delivery Failure

    EBI, LLC recalls 829 replacement lead wire assemblies (Part 1067724-2) distributed nationwide since May 1, 2023. Damaged wire sheaths may prevent therapeutic treatment signals from reaching electrodes, potentially delaying treatment.

    Product
    20" Lead Wires, Replacement Part Number: 1067724-2
    Category
    Distribution
    0 states