Siemens Medical Solutions USA, Inc. recalled 3 products on 2024-05-01

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-05-01
Critical
0
Units
5,785

Why these products were recalled

On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2024-05-01

    Diagnostic Ultrasound System Displays Incorrect Cardiac Values During Measurement Export

    ACUSON Maple ultrasound systems display the last cardiac measurement value instead of the configured minimum or maximum, potentially causing patient misdiagnosis during cardiac assessment.

    Product
    ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A
    Category
    Distribution
    44 states
  • HighFDA (Devices)··2024-05-01

    ACUSON Redwood 1.0 ultrasound displays incorrect cardiac measurement values

    The ACUSON Redwood 1.0 Diagnostic Ultrasound System may display the last measured cardiac value instead of the minimum or maximum value as configured, potentially leading to misdiagnosis or improper patient management.

    Product
    ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G
    Category
    Distribution
    44 states
  • HighFDA (Devices)··2024-05-01

    ACUSON Juniper ultrasound systems display incorrect cardiac measurements

    Siemens ACUSON Juniper ultrasound systems may display the wrong cardiac measurement value when configured to show minimum or maximum values, potentially leading to misdiagnosis.

    Product
    ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB10A to VB10G, VB11A, VB11B, VB30B and VB30C
    Category
    Distribution
    44 states