FDA (Devices) recall action 94138

Randox Laboratories Ltd. recalled 2 products on 2024-04-24

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-04-24
Critical
0
Units
14

Why these products were recalled

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-04-24

    Microalbumin Calibrator Series Recall Due to Measurement Accuracy Issue

    FDA recalls Microalbumin Calibrator Series due to negative measurement bias that may misclassify patient results. Laboratories using affected batches should review quality control data and consider recalibration.

    Product
    Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
    Category
    Distribution
    4 states