The Recall Desk

FDA (Devices) recall action 94128

Covidien, LP recalled 2 products on 2024-04-10

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-04-10
Critical
0
Units
3,053

Why these products were recalled

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2024-04-10

    Signia Surgical Stapler Reload recalled for uncontrolled articulation

    Covidien is recalling Signia Small Diameter Curved Tip Intelligent Reload surgical stapler reloads due to potential malfunction that could disrupt the staple line during use. The affected reloads may articulate uncontrollably due to unsecured components.

    Product
    Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-04-10

    Covidien Signia Surgical Stapler Reloads Recalled Due to Articulation Defect

    Covidien is recalling certain Signia surgical stapler reloads due to unsecured components that can cause uncontrolled articulation during use, potentially disrupting the staple line. The recall affects 3,023 units distributed worldwide.

    Product
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
    Category
    Distribution
    0 states