The Recall Desk

FDA (Devices) recall action 94090

FUJIFILM Healthcare Americas Corporation recalled 6 products on 2024-03-13

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2024-03-13
Critical
0
Units

Why these products were recalled

Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2024-03-13

    FUJIFILM Synapse PACS Software Produces Incorrect Breast Imaging Measurements

    FUJIFILM is recalling Synapse PACS Version 7.2.100 because the medical imaging software produces incorrect measurements when analyzing certain breast images. Affected healthcare facilities using this software version should verify measurement accuracy in their systems.

    Product
    Synapse PACS - Version 7.2.100
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-03-13

    Medical imaging software measurements incorrect in specific cases

    FUJIFILM's Synapse PACS Version 7.2.200 produces incorrect measurements when analyzing certain breast imaging scans without pixel spacing data. This measurement error could affect diagnostic accuracy.

    Product
    Synapse PACS - Version 7.2.200
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-03-13

    Medical Device Imaging Software May Calculate Incorrect Breast Measurements

    Synapse PACS Version 7.1.000US may produce incorrect measurements when Secondary Capture 2D images lack pixel spacing information in the DICOM header. The affected software is installed in healthcare facilities worldwide, including the United States and Puerto Rico.

    Product
    Synapse PACS - Version 7.1.000US
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-03-13

    Synapse PACS version 7.3.000 produces incorrect breast imaging measurements

    FUJIFILM's Synapse PACS version 7.3.000 software produces incorrect measurements when secondary capture images without pixel spacing are combined with breast tomosynthesis series. No illnesses or injuries reported.

    Product
    Synapse PACS - Version 7.3.000
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2024-03-13

    Synapse PACS Software Version 7.2.000 Measurement Error in Breast Imaging

    FUJIFILM's Synapse PACS Version 7.2.000 produces incorrect measurements in breast imaging when analyzing Secondary Capture 2D images lacking pixel spacing information. Affected healthcare facilities should discontinue use for breast imaging measurements.

    Product
    Synapse PACS - Version 7.2.000
    Category
    Distribution
    1 state