FDA (Devices) recall action 93939

Siemens Healthcare Diagnostics, Inc. recalled 2 products on 2024-03-20

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-03-20
Critical
0
Units

Why these products were recalled

Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-03-20

    Siemens Atellica IM Erythropoietin assay recalled for calibration bias error

    Siemens has recalled the Atellica IM Erythropoietin assay due to a 35% negative bias compared to WHO standards. The assay does not achieve claimed reference intervals and could report inaccurate test results.

    Product
    Atellica IM Erythropoietin (EPO) Assay (100 Test)
    Category
    Distribution
    12 states
  • HighFDA (Devices)··2024-03-20

    ADVIA Centaur Erythropoietin Assay Affected by -35% Measurement Bias

    Siemens Healthcare's ADVIA Centaur EPO Assay (lots 53984040, 55443042) produces readings 35% lower than the WHO standard, failing to achieve its claimed reference interval. No illnesses reported, but diagnostic inaccuracy poses risk of wrong clinical decisions.

    Product
    ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
    Category
    Distribution
    12 states