FDA (Devices) recall action 93927

Ortho-Clinical Diagnostics, Inc. recalled 5 products on 2024-03-13

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2024-03-13
Critical
0
Units

Why these products were recalled

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

The agency’s own wording, quoted from the enforcement report.

1–5 of 5

  • HighFDA (Devices)··2024-03-13

    Laboratory chemistry system software defect impairs quality control reporting

    A software defect in VITROS XT 3400 Chemistry Systems prevents quality control rules from properly reporting failures in test controls. This may result in erroneous patient test results being reported.

    Product
    VITROS XT 3400 Chemistry System, Catalog No. 6844458
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-03-13

    VITROS 5600 Diagnostic System Software Defect Prevents Quality Control Flagging

    Ortho-Clinical Diagnostics is recalling VITROS 5600 diagnostic systems (software versions 3.8.0 or 3.8.1) due to a defect that prevents quality control alerts from displaying, potentially allowing erroneous patient test results to be reported.

    Product
    VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-03-13

    VITROS Diagnostic System Software Defect May Cause Erroneous Patient Results

    Ortho-Clinical Diagnostics is recalling VITROS XT 7600 systems with software defects in versions 3.8.0 and 3.8.1. The defect prevents quality control rules from being properly reported, which could cause erroneous patient test results.

    Product
    VITROS XT 7600 Integrated System, Catalog No. 6844461
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-03-13

    VITROS 4600 Chemistry System Software Defect Affects Quality Control Reporting

    A software defect in VITROS 4600 Chemistry Systems running software versions 3.8.0 or 3.8.1 prevents quality control rules from being flagged or reported correctly, potentially allowing erroneous patient results to be released.

    Product
    VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-03-13

    Medical Device Software Defect May Cause Erroneous Lab Results

    A software defect in VITROS 3600 Immunodiagnostic Systems (versions 3.8.0 and 3.8.1) prevents quality control rules from reporting failures as expected, potentially allowing invalid test results to be reported.

    Product
    VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
    Category
    Distribution
    0 states