The Recall Desk

FDA (Devices) recall action 93788

Philips North America recalled 6 products on 2024-01-31

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2024-01-31
Critical
0
Units

Why these products were recalled

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–6 of 6

  • ModerateFDA (Devices)··2024-01-31

    MRI Cooling Cabinet Has Incorrect Electrical Warning Label

    Cooling cabinets in Philips MRI systems have defective electrical warning labels with incorrect colors and a missing high voltage warning symbol, creating electrical shock risk for service personnel.

    Product
    SmartPath to dStream for 1.5T, Model No. 781260, 782112
    Category
    Distribution
    7 states
  • ModerateFDA (Devices)··2024-01-31

    Philips MRI system recalled for incorrect electrical warning label

    Philips Achieva 3.0T MRI systems have an incorrect electrical warning label on the liquid cooling cabinet, missing the high voltage symbol and proper yellow background. This presents a risk of electrical shock to service personnel.

    Product
    Achieva 3.0T, Model No. 781177
    Category
    Distribution
    7 states
  • ModerateFDA (Devices)··2024-01-31

    Medical Device MRI Cooling Cabinet Warning Label Missing High-Voltage Symbol

    Philips Ingenia Ambition X MRI systems contain an incorrect warning label on the cooling cabinet. The label is missing high-voltage warning symbols, creating a potential electrical shock risk for service personnel.

    Product
    Ingenia Ambition X, Model No. 782109
    Category
    Distribution
    7 states