Philips North America recalled 12 products on 2024-02-07

The FDA (Devices) publishes a recall like this one as 12 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 12 pages.

Products
12
Announced
2024-02-07
Critical
0
Units

Why these products were recalled

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

The agency’s own wording, quoted from the enforcement report.

1–12 of 12

  • HighFDA (Devices)··2024-02-07

    Philips MRI Coil Seal Adhesive May Fail and Create Sharp Patient Contact Surfaces

    Philips has recalled certain SmartPath to dStream MRI coil components because the seal adhesive may fail, creating sharp edges that could contact patients during scanning and cause skin and tissue injuries.

    Product
    SmartPath to dStream for 1.5T. Model (REF) Numbers 781260, 782112.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    MRI Machine Coil Seal Adhesive May Create Sharp Edges

    The Intera Achieva 1.5T Pulsar MRI machine's Quadrature Body Coil seal adhesive may fail, creating sharp edges that could injure patients. All units of this model are affected.

    Product
    Intera Achieva 1.5T Pulsar. Model (REF) Number 781171.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Ingenia 3.0T CX MRI: Loose Coil Seal May Cause Sharp Edges

    Philips is recalling Ingenia 3.0T CX MRI machines due to potential failure of the Quadrature Body Coil seal adhesive. Loose seals may create sharp edges that risk skin injury or hair entanglement during scanning.

    Product
    Ingenia 3.0T CX. Model (REF) Numbers 781271, 782105.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Achieva MRI Scanner Recall: Loose Coil Seals May Cause Skin Injury

    Philips is recalling Achieva 1.5T MRI scanners due to a seal adhesive defect on the body coil that may fail during scanning, creating sharp edges that risk skin injury including abrasions and lacerations.

    Product
    Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Achieva XR MRI Coil Seal Adhesive Failure Risk

    Philips Achieva XR MRI body coil seal adhesive may fail and create sharp edges that could injure patients. Affected units are being recalled.

    Product
    Achieva XR. Model (REF) Numbers 781153, 781253.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Achieva 1.5T MRI system recalled for defective coil seal

    Philips is recalling the Achieva 1.5T Initial MRI system due to a defective Quadrature Body Coil seal. The adhesive may fail and create sharp edges that could cut or abrade patients during scanning.

    Product
    Achieva 1.5T Initial system. Model (REF) Numbers 781178.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Achieva 3.0T MRI Coil Seal Adhesive May Fail, Risking Patient Injury

    The Quadrature Body Coil seal adhesive in Philips Achieva 3.0T MRI scanners may fail, creating sharp edges that can injure patients. All units with model numbers 781277, 781177, 781278, 781344, and 781345 are affected.

    Product
    Achieva 3.0T. Model. (REF) Numbers 781277, 781177, 781278, 781344, 781345.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Philips Ingenia MRI Scanner Body Coil Seal Adhesive May Fail

    Philips is recalling Ingenia 1.5T CX MRI scanners because the Quadrature Body Coil seal adhesive may fail, creating sharp edges that could injure patients. Model numbers 781262 and 781261 are affected.

    Product
    Ingenia 1.5T CX. Model (REF) Numbers 781262, 781261.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    Achieva 1.5T MRI Coil Seal Adhesive May Fail Creating Sharp Edges

    Philips Achieva 1.5T MRI scanner components may experience coil seal adhesive failure, creating sharp edges that could injure patients. Affected devices have been distributed worldwide.

    Product
    Achieva 1.5T Conversion. Model (REF) Numbers 781346, 781283.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-02-07

    MRI Scanner Quadrature Coil Seal Failure Poses Patient Contact Risk

    Philips is recalling the Intera 1.5T Achieva Nova-Dual MRI scanner due to potential failure of the Quadrature Body Coil seal adhesive. If the seal fails, sharp edges may contact patients, risking skin injuries.

    Product
    Intera 1.5T Achieva Nova-Dual. Model (REF) Number 781173.
    Category
    Distribution
    0 states