FDA (Devices) recall action 93780

Wipro GE Healthcare Private Ltd. recalled 6 products on 2024-02-21

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2024-02-21
Critical
0
Units
26,329

Why these products were recalled

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–6 of 6

  • SevereFDA (Devices)··2024-02-21

    GE Healthcare Care Plus Incubator Hood Latching Defect Increases Fall Risk

    GE Healthcare incubator hoods can appear closed but may not be properly latched, posing a fall risk if patients impact an unsecured panel.

    Product
    GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2024-02-21

    GE Healthcare Incubators Bedside Panel Latch Failure Poses Fall Risk

    GE Healthcare incubators have defective bedside panels that may appear closed but not be properly latched, creating a patient fall risk when the panels disengage upon impact.

    Product
    GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2024-02-21

    GE Healthcare neonatal incubator bedside panels may fail to latch properly

    GE Healthcare is recalling neonatal incubators because the bedside panel can appear closed while unlatched. Impact could cause it to fall open, risking patient injury.

    Product
    GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2024-02-21

    GE Healthcare incubator bedside panel may disengage and fall open

    The bedside panel on GE Healthcare Care Plus incubators may appear closed while unlatched. If impacted, the panel can fall open, exposing patients to injury risk.

    Product
    GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2024-02-21

    GE Healthcare Lullaby Incubator bedside panels may fail to latch properly

    GE Healthcare Lullaby Incubator bedside panels and portholes may appear closed but not be properly latched. If impacted, these panels can disengage and fall open, creating a fall risk for patients.

    Product
    GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2024-02-21

    GE Healthcare Care Plus and Lullaby incubators recalled for bedside panel latching defect

    GE Healthcare is recalling 1,410 neonatal incubators worldwide because bedside panels and portholes can appear closed while not properly latched. If impacted, they may fall open and fail to protect patients from falls.

    Product
    GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR
    Category
    Distribution
    0 states