The Recall Desk

FDA (Devices) recall action 93771

Smiths Medical ASD Inc. recalled 4 products on 2024-04-24

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-04-24
Critical
0
Units
12,850

Why these products were recalled

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2024-04-24

    Smiths Medical syringe pumps recalled for molding defect in barrel clamp guide

    Smiths Medical is recalling certain Model 3010 syringe pumps due to a molding defect in the barrel clamp guide that may prevent proper syringe recognition. Affected units were manufactured or serviced between July 2016 and April 2021.

    Product
    Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-04-24

    Smiths Medical Medfusion Pump Guide Barrel Clamp Molding Defect Recall

    Smiths Medical is recalling Medfusion Guide Barrel Clamps due to a molding defect that could cause spring slippage, preventing the pump from recognizing or correctly identifying loaded syringes.

    Product
    Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-04-24

    Smiths Medical Medfusion syringe pumps may fail to recognize syringes

    Smiths Medical syringe pumps with defective barrel clamp guides may fail to recognize or correctly identify loaded syringes, potentially affecting medication delivery accuracy.

    Product
    Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-04-24

    Smiths Medical Medfusion Syringe Pumps May Fail to Recognize Syringes

    Smiths Medical is recalling certain Medfusion Models 4000, 3500, and 3010 syringe pumps due to a barrel clamp guide defect. The defect may prevent the pump from recognizing or correctly identifying a syringe, potentially affecting medication delivery.

    Product
    Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o)…
    Category
    Distribution
    0 states