The Recall Desk

FDA (Devices) recall action 93737

Philips North America recalled 3 products on 2024-02-14

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-02-14
Critical
0
Units
13

Why these products were recalled

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2024-02-14

    Ingenia 1.5T Patient Trolley Wheels May Attract to MRI Magnetic Field

    Philips is recalling Ingenia 1.5T patient transportation trolleys with magnetic wheels that could be attracted to MRI magnetic fields. These wheels pose a potential hazard in MRI-equipped facilities when the trolley operates near the magnetic field.

    Product
    Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2024-02-14

    Patient transportation system wheels may be attracted to MRI magnetic fields

    Ingenia Ambition X patient transportation system trolleys contain magnetic wheels that may be attracted to MRI magnetic fields, potentially causing trolley malfunction during patient procedures.

    Product
    Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2024-02-14

    Patient Transportation System Recalled for Magnetic Wheel MRI Interference Risk

    Philips is recalling Ingenia Elition X patient transportation systems due to magnetic wheels that may be attracted to MRI systems, posing a patient safety risk if used in MRI environments.

    Product
    Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
    Category
    Distribution
    2 states