The Recall Desk

FDA (Devices) recall action 93655

Beckman Coulter, Inc. recalled 4 products on 2024-01-31

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2024-01-31
Critical
0
Units
384

Why these products were recalled

Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2024-01-31

    iQ Waste Well Adapter Recalled for Rust in Cannula Component

    Beckman Coulter recalls the iQ Waste Well Adapter due to rust in the cannula that may generate erroneous test results in laboratory diagnostic equipment.

    Product
    iQ Waste Well Adapter, Part Number: 700-3393
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-31

    Beckman Coulter iQ200 Analyzers Recalled for Potential Rust and Diagnostic Errors

    Beckman Coulter is recalling 112 units of iQ200 Series Urine Microscopy Analyzers distributed globally. The cannula in the Flowcell and Rinse Waste Well Assembly may rust, potentially causing erroneous diagnostic results.

    Product
    iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-31

    Beckman Coulter DxU Analyzers Recalled for Potential Rust-Related Diagnostic Errors

    Beckman Coulter is recalling DxU 850m/840m Microscopy Analyzers (254 units) due to potential rust in the cannula component, which may cause erroneous test results. The recall affects units distributed in the US and multiple international locations.

    Product
    Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
    Category
    Distribution
    0 states