Why these products were recalled

The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2024-01-24

    Medical Fiber Cleaver System Recalled for Contamination Risk

    Olympus Corporation recalled 236 fiber cleaver units distributed non-sterile but labeled as reusable and autoclavable. Without validated sterilization instructions, use on sterile fiber poses contamination risk.

    Product
    Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Olympus Laser Fiber Stripper Recall Due to Sterilization Validation Gap

    Olympus recalls fiber stripper models sold before 2023 lacking validated sterilization instructions, creating a contamination risk when used on sterile medical fibers.

    Product
    Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940
    Category
    Distribution
    0 states