FDA (Devices) recall action 93577

Icotec Ag recalled 2 products on 2024-01-17

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-01-17
Critical
0
Units
79

Why these products were recalled

icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2024-01-17

    VADER Pedicle System Torque Wrench Recalled Due to Defective Tightening Function

    Icotec is recalling VADER pedicle system torque wrenches because the torque limiting function can malfunction, potentially leaving screw fasteners insufficiently tightened during implant surgery.

    Product
    VADER pedicle system, torque wrench, Catalog Number 42-703
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2024-01-17

    VADER pedicle system torque wrench may not tighten implant screws correctly

    Icotec Ag is recalling VADER pedicle system torque wrenches (Catalog Number 42-702) because the torque limiting function may malfunction, potentially leaving nut screws insufficiently tightened during implant surgery. No injuries have been reported.

    Product
    VADER pedicle system, torque wrench, Catalog Number 42-702
    Category
    Distribution
    10 states