FDA (Devices) recall action 93506

Smiths Medical Asd Inc recalled 9 products on 2024-01-24

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2024-01-24
Critical
0
Units
285,165

Why these products were recalled

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–9 of 9

  • HighFDA (Devices)··2024-01-24

    FDA Recalls Medical Pressure Tubing With Defective O-Ring Seals

    Smiths Medical is recalling 31,685 units of pressure tubing with defective O-rings that may not seal properly during medication infusion, potentially causing medication leaks and underdosing. The defect affects specific lot numbers with worldwide distribution.

    Product
    6IN PRESSURE TUBING, List Number MX20617
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Medex Injection Line Rotating Adapter Recalled Due to O-Ring Defect

    Smiths Medical is recalling Medex injection line rotating adapters due to a manufacturing defect where the O-ring inner diameter may be oversized, potentially causing medication leaks during infusion.

    Product
    Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Medex LOGICAL CATH LAB KIT Recalled for Oversized O-Ring Defect

    Smiths Medical is recalling the Medex LOGICAL CATH LAB KIT due to a manufacturing defect where the O-ring inner diameter may be oversized, affecting seal integrity and potentially leading to medication under-infusion.

    Product
    Medex LOGICAL CATH LAB KIT, List Number M20754
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Medex 700PSI Stopcock with Rotator Recalled Due to Seal Defect

    Smiths Medical is recalling Medex 700PSI stopcocks with rotators due to a manufacturing defect affecting the O-ring seal. The defect may cause medication leakage and under-infusion during patient treatment.

    Product
    Medex 700PSI STOPCOCK W/ROTATOR, List Number MX4331R
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Medical Device Stopcock Recalled Due to O-Ring Seal Defect

    Smiths Medical Asd Inc is recalling approximately 31,685 Stopcock Left Rotator devices due to oversized O-rings that may compromise seal integrity. The defect may cause medication leaks, potentially resulting in under-infusion during treatment.

    Product
    1050 STOPCOCK LEFT ROTATOR OFF HANDLE, List Number MX1431MRL
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Smiths Medical Injection Line Rotating Adapters Recalled for Seal Defect

    Smiths Medical is recalling specific injection line rotating adapters due to a manufacturing defect causing oversized O-rings that may lead to inadequate seal integrity and medication under-infusion.

    Product
    48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    High Pressure Rotator Infusion Device Recalled Due to Seal Defect

    Smiths Medical is recalling 31,685 HIGH PRESSURE ROTATOR infusion devices with a manufacturing defect that may cause an inadequate seal. The defect could lead to medication leaks and under-infusion.

    Product
    HIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HP
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    High Pressure Rotator with O-Ring Seal Defect Recalled Worldwide

    Smiths Medical is recalling High Pressure Rotators due to a manufacturing defect affecting O-ring seals. An oversized inner diameter may cause inadequate seal integrity, risking medication under-infusion.

    Product
    HIGH PRESSURE ROTATOR WITH FEMALE LUER LOCK, List Number MX497HP
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-24

    Braided Injection Line Recalled for Potential Medication Leaks

    Smiths Medical is recalling approximately 31,685 units of a braided injection line due to manufacturing defects in the O-ring seal. The defect may cause medication leaks that reduce drug delivery.

    Product
    20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BR
    Category
    Distribution
    0 states