The Recall Desk

FDA (Devices) recall action 93499

ResMed Ltd. recalled 7 products on 2023-12-27

The FDA (Devices) publishes a recall like this one as 7 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 7 pages.

Products
7
Announced
2023-12-27
Critical
0
Units
28,748,349

Why these products were recalled

Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–7 of 7

  • SevereFDA (Devices)··2023-12-27

    AirTouch N20 Nasal Mask recalled for magnetic interference with medical implants

    ResMed is recalling 424,309 units of AirTouch N20 Nasal Masks because they contain magnets that may interfere with active medical implants and ferromagnetic materials. Updated safety warnings are being provided.

    Product
    AirTouch N20 Nasal Mask and User Guide
    Category
    Distribution
    16 states
  • SevereFDA (Devices)··2023-12-27

    AirFit F30i Full Face Mask Recalled Due to Magnetic Interference Risk

    ResMed is recalling the AirFit F30i Full Face Mask due to potential magnetic interference with active medical implants and ferromagnetic objects. Safety warnings are being updated.

    Product
    AirFit F30i Full Face Mask and User Guide
    Category
    Distribution
    16 states
  • SevereFDA (Devices)··2023-12-27

    AirFit N20 Nasal Mask: Magnetic Interference with Medical Implants

    ResMed AirFit N20 Nasal Masks contain magnets that may interfere with active medical implants and ferromagnetic objects. The FDA classified this as Class I; ResMed is updating safety warnings with required distances.

    Product
    AirFit N20 Nasal Mask and User Guide
    Category
    Distribution
    16 states
  • SevereFDA (Devices)··2023-12-27

    AirFit F30 Full Face Mask recalled for magnetic interference with medical implants

    ResMed is recalling the AirFit F30 Full Face Mask because the device contains magnets that can interfere with active medical implants and ferromagnetic objects. The original user guide did not adequately warn about safe distance requirements.

    Product
    AirFit F30 Full Face Mask and User Guide
    Category
    Distribution
    16 states