The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2024-01-03
Critical
0
Units
176,843

Why these products were recalled

Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2024-01-03

    Medtronic SynchroMed II Pump May Fail to Resume Drug Delivery After MRI

    The Medtronic SynchroMed II pump may fail to resume drug delivery after MRI if electromagnetic interference triggers telemetry mode while the pump is alarming. This could interrupt essential medication delivery in patients.

    Product
    Medtronic SynchroMed II, Model 8637-20, Programmable pump
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-03

    Medtronic SynchroMed II Programmable Pump May Fail to Resume Drug Delivery After MRI

    The Medtronic SynchroMed II pump may fail to resume medication delivery after an MRI scan if the device switches to telemetry mode due to electromagnetic interference while alarming. The pump will not resume normal operation after leaving the MRI magnetic field.