The Recall Desk

FDA (Devices) recall action 93479

GE Hangwei Medical Systems, Co. LTD recalled 2 products on 2023-12-20

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-12-20
Critical
0
Units
202

Why these products were recalled

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-12-20

    X-ray system overhead suspension components may fall due to incorrect bolt torque

    GE HealthCare Definium Tempo and Tempo Pro X-ray systems may have been assembled with incorrect bolt torque, creating a risk that Overhead Tube Suspension components could fall.

    Product
    The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-20

    GE Definium Tempo X-Ray System Bolts May Loosen and Fall

    Incorrect bolt torque on GE Definium Tempo X-ray systems may cause the overhead tube suspension assembly to fall. No injuries have been reported, but GE recommends immediate inspection of affected systems.

    Product
    The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD…
    Category
    Distribution
    0 states