FDA (Devices) recall action 93473

Siemens Healthcare Diagnostics Inc recalled 4 products on 2023-12-27

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-12-27
Critical
0
Units
16,151

Why these products were recalled

RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2023-12-27

    RAPIDPoint 500 Measurement Cartridge Sodium Sensor May Cause Measurement Error

    Siemens RAPIDPoint 500 cartridges may produce inaccurate sodium measurements in blood samples, potentially delaying diagnosis or causing inappropriate treatment. The sodium sensor may show a negative bias, affecting patient care.

    Product
    RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491448
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-12-27

    RAPIDPoint 500 Diagnostic Cartridges Recall Due to Sodium Sensor Measurement Bias

    Siemens is recalling RAPIDPoint 500 measurement cartridges because the sodium sensor can produce inaccurate readings, potentially leading to delayed diagnosis or incorrect treatment of electrolyte imbalances.

    Product
    RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491447
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-12-27

    RAPIDPoint 500 Measurement Cartridge Sodium Sensor Negative Bias Risk

    Siemens RAPIDPoint 500 measurement cartridges may produce inaccurate sodium readings, risking delayed diagnosis of dangerous electrolyte imbalances or unnecessary treatment.

    Product
    RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491449
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-12-27

    RAPIDPoint 500 Measurement Cartridges Recalled for Sodium Sensor Bias

    Siemens is recalling RAPIDPoint 500 Systems Measurement Cartridges because the sodium sensor can develop a negative measurement bias that could delay diagnosis of serious electrolyte disorders or cause treatment errors.

    Product
    RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10844813
    Category
    Distribution
    1 state