The Recall Desk

FDA (Devices) recall action 93347

CardioQuip, LLC recalled 2 products on 2023-12-13

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-12-13
Critical
0
Units
275

Why these products were recalled

There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2023-12-13

    CardioQuip Modular Cooler-Heater Units Recalled for Heating Chamber Malfunction

    CardioQuip Modular Cooler-Heater units are being recalled due to a potential heating-chamber malfunction that could cause melting and burning of internal components. 125 units distributed across 15 US states are affected.

    Product
    CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A
    Category
    Distribution
    15 states
  • HighFDA (Devices)··2023-12-13

    CardioQuip Modular Cooler-Heater heating chamber malfunction poses burn risk

    CardioQuip is recalling 150 units of its Modular Cooler-Heater (Model MCH-1000m) due to potential heating-chamber malfunction that could cause component melting or burning. The device may overheat unexpectedly.

    Product
    CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A
    Category
    Distribution
    15 states