The Recall Desk

FDA (Devices) recall action 93335

Aizu Olympus Co., Ltd. recalled 5 products on 2023-12-06

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2023-12-06
Critical
0
Units
120

Why these products were recalled

Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–5 of 5

  • HighFDA (Devices)··2023-12-06

    Improperly Repaired Gastrointestinal Videoscope May Expose Patients to Contamination

    Aizu Olympus gastrointestinal videoscopes (GIF-H190) may have defective repairs that prevent proper sterilization, risking patient exposure to contaminated material. Affected units were distributed to medical facilities in six states.

    Product
    GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
    Category
    Distribution
    6 states
  • HighFDA (Devices)··2023-12-06

    Olympus colonoscope repair defect may impair imaging and suction function

    Aizu Olympus is recalling 24 colonoscopes with defective repairs to the light guide connector. The defect may prevent proper imaging, reduce suction, and impair device cleaning, potentially exposing patients to contamination.

    Product
    CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
    Category
    Distribution
    6 states
  • HighFDA (Devices)··2023-12-06

    Gastrointestinal Videoscope Recall Due to Improper Repair and Infection Risk

    Aizu Olympus recalls 24 gastrointestinal videoscopes with improperly repaired light guide connectors. The defect may cause loss of image and suction, preventing proper device reprocessing and risking patient infection.

    Product
    GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
    Category
    Distribution
    6 states
  • HighFDA (Devices)··2023-12-06

    Olympus EVIS EXERA III Colonoscope Recall for Connector Defect

    24 Olympus EVIS EXERA III colonoscopes distributed to six states are recalled due to improperly repaired connectors that may cause image loss, decreased suction, or improper reprocessing, potentially exposing patients to contaminated equipment.

    Product
    PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
    Category
    Distribution
    6 states