FDA (Devices) recall action 93313

Instrumentation Laboratory recalled 3 products on 2023-11-29

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2023-11-29
Critical
0
Units
7,344

Why these products were recalled

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2023-11-29

    Medical Device Cartridges Recalled for Blood Oxygen Reading Errors

    Instrumentation Laboratory is recalling GEM Premier 3000 PAK cartridges due to a defect that can produce falsely high blood oxygen readings. The inaccuracy may prevent detection of dangerously low oxygen levels and delay necessary treatment.

    Product
    GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2023-11-29

    Blood Gas Analyzer Cartridges Recalled for Falsely Elevated Oxygen Readings

    Instrumentation Laboratory recalls GEM Premier 3000 PAK blood gas analyzer cartridges due to measurement errors in low oxygen detection. The devices may falsely indicate normal oxygen levels when levels are dangerously low, potentially delaying treatment.

    Product
    GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2023-11-29

    Blood Gas Analyzer Cartridges Recalled for Potential Oxygen Measurement Errors

    The FDA is recalling GEM Premier 3000 PAK cartridges manufactured by Instrumentation Laboratory because they may incorrectly report high oxygen levels in patients with low blood oxygen, potentially delaying critical treatment.

    Product
    GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
    Category
    Distribution
    3 states