Philips North America Llc recalled 3 products on 2023-11-22

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2023-11-22
Critical
0
Units
96

Why these products were recalled

For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to in…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2023-11-22

    Philips Brilliance CT Big Bore May Send Incorrect Coordinates During Radiation Therapy Planning

    Philips Brilliance CT Big Bore units with software V4.8.0.10421 may produce inaccurate image position data when offset reconstruction is used. This could lead to incorrect radiation therapy planning or diagnosis.

    Product
    Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and therapy.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-22

    Philips Big Bore RT CT Scanner Defect May Cause Incorrect Treatment Planning

    A software defect in Philips Big Bore RT CT scanners may cause incorrect coordinates in radiation therapy planning systems, affecting 56 units distributed nationwide and internationally.

    Product
    Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation oncology and therapy.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-22

    CT Scanner Image Fusion Software Error Affecting Radiation Therapy Planning

    A software bug in Philips Brilliance CT scanners causes image-fusion errors, resulting in incorrect radiation therapy coordinates and potential treatment planning failures.

    Product
    Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator designed for radiation oncology and therapy.
    Category
    Distribution
    0 states