The Recall Desk

FDA (Devices) recall action 93269

Unomedical A/S recalled 3 products on 2023-11-29

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2023-11-29
Critical
1
Units
13,069

Why these products were recalled

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • SevereFDA (Devices)··2023-11-29

    VariSoft Infusion Set Model 1002828 Recalled for Connector Detachment Risk

    The FDA recalls 1,679 VariSoft infusion sets (Model 1002828) due to connector damage that causes improper detachment. Undetected disconnections can lead to missed insulin doses and diabetic ketoacidosis.

    Product
    VariSoft infusion set, Model Number 1002828
    Category
    Distribution
    1 state
  • CriticalFDA (Devices)··2023-11-29

    VariSoft Insulin Infusion Set Recalled Over Premature Connector Disconnection

    Unomedical recalls 3,362 units of VariSoft infusion sets due to connector damage causing premature disconnection and missed insulin doses, risking dangerous elevation of blood glucose and ketone levels.

    Product
    VariSoft infusion set, Model Number 1002830
    Category
    Distribution
    1 state