The Recall Desk

FDA (Devices) recall action 93264

Elekta, Inc. recalled 6 products on 2023-12-13

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2023-12-13
Critical
0
Units
2,387

Why these products were recalled

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–6 of 6

  • HighFDA (Devices)··2023-12-13

    Medical Radiation Therapy System Detector Arm Malfunction Poses Uncontrolled Extension Risk

    Elekta recalled its ELEKTA AXESSE radiation therapy system due to potential uncontrolled extension of detector arms if the microswitch is incorrectly configured. The recall affects 47 units distributed worldwide.

    Product
    ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-13

    Medical Radiation Therapy Device Microswitch May Cause Uncontrolled Detector Arm Extension

    Elekta radiation therapy accelerators may experience uncontrolled detector arm extension due to an incorrectly set microswitch in the middle arm. The issue affects 315 units worldwide.

    Product
    ELEKTA LIMITED Precise Digital Accelerator, REF MRT 6001, MRT 6011; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-13

    Medical radiation therapy system detector arm extension recall

    Elekta recalls Infinity radiation therapy systems for a detector arm defect that could cause uncontrolled extension if a microswitch is improperly configured. No injuries reported.

    Product
    ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-13

    Medical Radiation Therapy System Potential Detector Arm Malfunction Due to Microswitch Issue

    Recall of approximately 1,019 Elekta Synergy radiation therapy systems due to potential uncontrolled extension of detector arms if the middle arm microswitch is incorrectly setup.

    Product
    ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-13

    Radiation Therapy System Detector Arm Microswitch Defect May Cause Uncontrolled Extension

    Elekta is recalling 264 units of its Synergy Platform radiation therapy systems due to a microswitch defect that could cause uncontrolled extension of detector arms. The systems were distributed worldwide.

    Product
    ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-12-13

    Elekta Versa HD radiation therapy system detector arm uncontrolled extension risk

    Certain Elekta Versa HD radiation therapy systems may experience uncontrolled detector arm extension due to microswitch setup issues. This recall affects 451 systems worldwide.

    Product
    ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
    Category
    Distribution
    0 states