The Recall Desk

FDA (Devices) recall action 93237

SunMed Holdings, LLC recalled 9 products on 2023-11-29

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2023-11-29
Critical
0
Units
—

Why these products were recalled

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–9 of 9

  • HighFDA (Devices)··2023-11-29

    Medline Manual Resuscitators Recalled for Backwards Manometer Leak

    Medline manual resuscitators are recalled due to a backwards leak in the integrated manometer that allows CO2 rebreathing. The defect affects multiple models used in respiratory care distributed nationwide.

    Product
    Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    V-Care manual resuscitators recalled for CO2 rebreathing defect

    SunMed's Ventlab V-Care manual resuscitators are being recalled due to a backwards leak in the integrated manometer that allows CO2 rebreathing. Affected devices were distributed nationwide.

    Product
    Ventlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN4100MCP; 2. V-Care Infant Resuscitator REF #s: VN3100OB, VN3100MB-2, VN3100MB-M01, VNW3100MB-2…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    Manual resuscitators recalled due to carbon dioxide rebreathing defect

    SunMed Holdings recalls Ventlab STAT-Check manual resuscitators due to a backwards leak in the integrated manometer allowing CO2 rebreathing, which may impair ventilation. Multiple models (Adult, Infant, Pediatric) are affected and distributed nationwide.

    Product
    Ventlab, LLC STAT-Check manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. STAT-Check Adult Resuscitator REF #: SC9101B; 2. STAT-Check Infant Resuscitator RES #s: SC7101B & SC7101B-M0; 3. STAT-Check Pediatric Resuscitator…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    RescuMed Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings recalls RescuMed manual resuscitators due to a backwards leak in the manometer allowing CO2 rebreathing. Affected models distributed nationwide across 32 US states.

    Product
    Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3.…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    Horizon manual resuscitators recalled for CO2 rebreathing hazard

    SunMed Holdings is recalling Ventlab Horizon manual resuscitators due to a backwards leak in the integrated manometer that allows carbon dioxide rebreathing. The affected devices were manufactured from September 2018 to the present and distributed nationwide.

    Product
    Ventlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, 9106HZ, 9110HZ, 9120HZ, 9111HZ, 9100HZ, 9105HZ, 9108HZ, 9006HZ, 9002HZ; 2. Horizon Infant…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    Ventlab AirFlow Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings is recalling Ventlab AirFlow manual resuscitators nationwide due to a backwards leak in the integrated manometer that allows CO2 rebreathing. Users should immediately discontinue use and contact the manufacturer.

    Product
    Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    SafeSpot Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings is recalling SafeSpot manual resuscitators with integrated manometers due to a backwards leak that allows CO2 rebreathing. The defect affects multiple accessory configurations distributed nationwide.

    Product
    Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    Curaplex Manual Resuscitators Recalled for Potential CO2 Rebreathing Risk

    SunMed Holdings is recalling Curaplex manual resuscitators with integrated manometers due to a backwards leak in the patient valve that allows carbon dioxide rebreathing. All units manufactured from September 2018 to present and distributed nationwide are affected.

    Product
    Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer…
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2023-11-29

    BreathTech Manual Resuscitators Recalled for Integrated Manometer Leak

    SunMed Holdings is recalling BreathTech manual resuscitators because the integrated manometer can leak, allowing patients to rebreathe exhaled CO2 during emergency airway management.

    Product
    Ventlab, LLC BreathTech manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. BreathTech Adult Resuscitator REF #s: BT5107F & BT5104P; 2. BreathTech Child Resuscitator REF #s: BT3030, BT3034, & BT3025; 3. BreathTech Infant…
    Category
    Distribution
    31 states