FDA (Devices) recall action 93086

American Contract Systems, Inc. recalled 12 products on 2023-11-01

The FDA (Devices) publishes a recall like this one as 12 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 12 pages.

Products
12
Announced
2023-11-01
Critical
0
Units
2,355

Why these products were recalled

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–12 of 12

  • HighFDA (Devices)··2023-11-01

    IV catheter components recalled for improper sterilization exposure

    American Contract Systems is recalling CVS PCSU SJH IV catheter and adhesive components due to improper ethylene oxide sterilization. The components may have reduced functionality or elevated chemical residuals.

    Product
    CVS PCSU SJH, REF SJCV48J
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Cardiovascular Surgical Device Components Recalled for Improper Sterilization

    American Contract Systems recalls cardiovascular surgical device components that were subjected to improper ethylene oxide sterilization at excessive temperatures, potentially causing loss of functionality or drug efficacy.

    Product
    Vasc. Open CVOR SJH, REF SJVO56B
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Medical Device Packs Recalled for Improper Sterilization Process

    American Contract Systems recalled General Pack medical device components (I.V. catheters and adhesive) that were improperly sterilized with ethylene oxide at higher than approved temperatures. The components may lack functionality or have elevated sterilization residuals.

    Product
    General Pack, REF BBGP31C
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Pediatric Heart SJH Device Recall Due to Improper Sterilization

    American Contract Systems recalls Pediatric Heart SJH components exposed to improper sterilization, risking functionality loss and reduced drug efficacy.

    Product
    Pediatric Heart SJH, REF SJPH34K
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Robotics Pack I.V. Catheter and Adhesive Recalled for Sterilization Defect

    American Contract Systems is recalling its Robotics Pack (56 cases) due to improper sterilization of IV catheter and adhesive components. The devices may have lost functionality or drug efficacy and may contain excessive ethylene oxide residuals.

    Product
    Robotics Pack, REF SSRO22D
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Pediatric Urology Equipment Recalled Due to Improper Sterilization

    American Contract Systems is recalling Pediatric Urology Wolson Packs due to components being exposed to unapproved sterilization processes. Components may have lost functionality or reduced efficacy, potentially compromising device performance.

    Product
    Pediatric Urology Wolson Pack, REF BPPU55A
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2023-11-01

    Sub Q Recorder Pack components may lose functionality due to improper sterilization

    American Contract Systems is recalling Sub Q Recorder Pack components that were improperly sterilized using ethylene oxide at excessive temperatures. The I.V. Cath and Dermabond components may lose functionality or drug efficacy as a result.

    Product
    Sub Q Recorder Pack, REF BPRC87A
    Category
    Distribution
    2 states