3M Company - Health Care Business recalled 4 products on 2023-10-04

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-10-04
Critical
0
Units
362,748

Why these products were recalled

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-10-04

    3M Attest Biological Indicators Cap May Melt During Sterilization

    3M is recalling 77,280 Attest biological indicators due to caps that may melt, deform, or crack during steam sterilization cycles, potentially compromising process validation.

    Product
    3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-10-04

    3M Attest Biological Indicators Recalled for Defective Caps

    3M is recalling specific lots of Attest biological indicators used in steam sterilization monitoring. Certain lots have caps that may melt, deform, or crack at sterilization temperatures of 132°C and 135°C, potentially compromising sterilization process verification.

    Product
    3M Attest Super Rapid Readout Steam Challenge Pack, REF 1496V; designed to qualify and monitor dynamic-air removal steam sterilization processes
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-10-04

    3M Attest Sterilization Indicator Caps May Melt or Crack

    3M is recalling 172,464 units of Attest biological indicators used to validate steam sterilization because caps may melt, deform, or crack during sterilization at 132°C and 135°C.

    Product
    3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-10-04

    3M Attest Biological Indicator caps may fail during steam sterilization

    3M is recalling 79,500 Attest Super Rapid Readout Biological Indicators because the cap may melt, deform, or crack during steam sterilization cycles, potentially compromising sterilization validation.

    Product
    3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles
    Category
    Distribution
    0 states