Why these products were recalled

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2023-10-11

    Olympus SPiN Thoracic Navigation System Models Recalled for EMC Non-Conformance

    Olympus is recalling SPiN Thoracic Navigation System and related models due to manufacturing non-conformances and insufficient electromagnetic compatibility (EMC) evidence. The devices may malfunction in certain electrical environments.

    Product
    Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 0095 SYS-3000 SPiN Thoracic Navigation System SYS-240
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    Medical Navigation System Recalled for Manufacturing and Electromagnetic Compatibility Defects

    Olympus Veran percutaneous navigation systems are recalled due to manufacturing non-conformances and insufficient electromagnetic compatibility evidence, affecting approximately 9,961 units distributed worldwide.

    Product
    Veran : Percutaneous Models: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" INS-5611 SPiN Perc¿ Localization Needle - 2cm, 5/Bx INS-5610 SPiN Perc¿ Localization Needle - 1cm, 5/Bx INS-5600 SPiN Perc¿ Biopsy Needle Guide Kit, 5/Bx
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    Veran Endobronchial Navigation System: Manufacturing and EMC Defects

    Olympus is recalling Veran endobronchial navigation systems due to manufacturing non-conformances and insufficient electromagnetic compatibility evidence. Devices may malfunction or interfere with other medical equipment.

    Product
    Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    Veran Navigation System Biopsy Needle Devices: Electromagnetic Compatibility Issue

    Olympus is recalling 522 units of Veran navigation system biopsy needles due to manufacturing defects and insufficient electromagnetic compatibility testing. Devices may not function properly in certain electromagnetic environments.

    Product
    Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    Veran Endobronchial Navigation Devices Recalled for Design Non-Conformance and EMC Issues

    Olympus Corporation recalls Veran endobronchial navigation devices due to manufacturing non-conformances and insufficient electromagnetic compatibility. Approximately 6,318 units distributed worldwide are affected.

    Product
    Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend¿ 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC INS-5411 0.016" Stylet, 1235 mm L, 10/Bx INS-5410 SPiN Flex¿ ANSO¿ Needle, 22ga, 1.8mm OD, 5/Bx
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    Veran Endobronchial Navigation Devices Recalled for Manufacturing Non-conformance

    Olympus is recalling Veran endobronchial navigation devices (6,484 units) worldwide due to manufacturing non-conformances and insufficient evidence of adequate electromagnetic compatibility. The devices may not function acceptably in environments with electromagnetic interference.

    Product
    Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter¿ 90 for u
    Category
    Distribution